GX-Force Sample Draw 4 Gas Monitor

The GX-Force Sample Draw 4 Gas Monitor is RKI’s smallest personal 1-4 gas monitor with an internal sample pump. The GX-Force Sample Draw 4 Gas Monitor can monitor the standard confined space gases (LEL, Oxygen, Carbon Monoxide, and Hydrogen Sulfide). Utilising micro-sensor technology.

The Riken GX-Force has two operating modes, the GX-Force can be used for confined space, safety monitoring in its Normal Operating mode. A Leak Check mode is the solution for leak investigations. In Leak Check Mode, the GX-Force can monitor for combustibles in the ppm range.

The GX-Force runs for 30 hours on a rechargeable Li-ion battery, and has a large LCD display showing all gas readings, battery level, current time and automatically backlights in alarm conditions. Standard alarm types include vibration, visual, and audible alarms, which can be set to latching or non-latching. Controlled by a microprocessor, the GX-Force continuously checks itself for sensor connections, low battery, circuit trouble, low flow, and calibration errors.

Calibration and bump test intervals and reminders are user adjustable and can be set to either go into alarm or to lock the user out of normal measurement mode once a calibration period has expired. Calibration can be done individually or in a group.

GX-Force-Sample-Draw-4-Gas-Monitor

 

Sampling MethodInternal sample pump, flow rate nominal 0.5 LPM, includes hydrophobic filter
Suction Flow RateMinimum 0.35 L/min (open flow rate)
Gas Alarm PatternLamp flashing, continuous modulating buzzer sounding, gas concentration display flashing, vibration
Gas Alarm Reset OperationSelf-resetting or latching
Fault Alarm/Self-DiagnosisSystem, clock, or sensor abnormality; battery voltage drop; calibration failure;
pump abnormality; low flow rate
Fault Alarm PatternLamp flashing, intermittent buzzer sounding, detail display
Fault alarm reset operationSelf-resetting
DisplayLCD digital (7-segment + 14-segment + icons) with backlight
Individual OperationsOperational status, clock, battery level, peak reading, pump status, calibration notification
Sound PressureApprox. 90 dB (30 cm)
Data Logger FunctionMaximum storage capacity: 3,600 items Interval: 5 minutes (adjustable)
Communication SpecificationsUSB, 2.0 (for data logger) Connector: Type-C
Power SourceRechargeable lithium ion battery, USB Type-C power cord
Continuous Operating TimeApprox. 30 hours (25 °C, fully charged, no alarm, no backlight)
Operating Ambient
Temperature /Humidity Range
-20 °C to +50 °C (no sudden changes), 0 – 95 % RH (no condensation)
ApprovalsIntrinsically safe construction
IECEx (Ex da ia IIC T4 Ga/Ex ia IIC T4 Ga)
ATEX (II1G Ex da ia IIC T4 Ga/II1G Ex ia IIC T4 Ga)
Certifications

CQPSUS, Class I, Div 1, Groups ABCD

Ex da ia IIC T4 Ga (w/ NCR-6309 sensor)/Ex ia IIC T4 Ga (w/o NCR-6309 sensor)

Class I, Zone 0

AEx da ia IIC T4 Ga (w/ NCR-6309 sensor)/AEx ia IIC T4 GA (w/o NCR-6309 sensor)

Protection LevelIP67 equivalent
External Dimensions / WeightApprox. 2.52 (W) × 6.81 (H) × 1.85 (D) (excluding protrusions) / Approx. 9.87 oz.
WarrantyThree years material and workmanship

 

Detection target gas

Combustible gas
(CH4 or HC)

Oxygen
(O2)

Carbon monoxide
(CO)

Hydrogen sulfide
(H2S)

Detection principle

Catalytic combustion

Electrochemical type

Detection range

0 – 100 % LEL

0.0 – 40.0 vol %

0 – 2,000 ppm

0.0 – 200.0 ppm

Resolution

1 % LEL

0.1 vol %

1 ppm

0.1 ppm

Alarm setpoints

(User-defined setting)

1st alarm: 10 % LEL

2nd alarm: 25 % LEL

3rd alarm: 50 % LEL

OVER alarm: 100 % LEL

L alarm: 19.5 vol %

LL alarm: 18.0 vol %

H alarm: 23.5 vol %

OVER alarm: 40.0 vol %

1st alarm: 25 ppm

2nd alarm: 50 ppm

3rd alarm: 1200 ppm

TWA alarm: 25 ppm

STEL alarm: 200 ppm

OVER alarm: 2,000 ppm

1st alarm: 5.0 ppm

2nd alarm: 30.0 ppm

3rd alarm: 100.0 ppm

TWA alarm: 1.0 ppm

STEL alarm: 5.0 ppm

OVER alarm: 200.0 ppm

Response time (T90)

CH4: Within 30 seconds,

HC: Within 40 seconds

Within 20 seconds

Within 30 seconds

Within 30 seconds

View more information on Riken’s website